The LC‑I1100 / LC‑G1100 system offers a reliable and cost-effective HPLC platform for various analytical needs. Featuring a VFD display with both internal and external control modes. The P‑H1100 pump delivers high-precision flow from 0.001 to 10.000 mL/min with accuracy better than ±0.3% and precision RSD <0.1%. The UV‑VIS1100 detector covers 190–700 nm with noise ≤±1.0×10⁻⁵ AU and drift ≤2.0×10⁻⁴ AU/h. The column oven CTB‑1100 maintains temperature from RT + 5 °C to 99 °C with ±0.1 °C accuracy. Includes GM‑15 gradient mixer, Rheodyne 7725i injector, and a Supersil ODS2 (5 µm, 4.6×200 mm) analytical column. Comes with 18-month warranty, 24/7 support, and free accessories. Easy maintenance makes it ideal for food, pharmaceutical, environmental, and biotech labs.
GC‑900 Gas Chromatograph with FID/TCD/FPD Detectors
0 UZS
Found this product cheaper?
Specifications:
(see all)
Measured parameters
Separation, identification, and quantification of dissolved compounds
Measurement range
Detector: 190–700 nm
Measurement resolution
Flow rate resolution: 0.001 mL/min
Measurement accuracy
Pump accuracy ±0.3%, oven temp accuracy ±0.1 °C
Detector / Detector type
UV-VIS1100 UV-Vis detector